Technical Specification:
1. Power supply: AC 220V±10% 50Hz
2. Power: 500VAwith dental chair
3. The leak current to the ground: ≤0.5mA
4. Resistance insulation: ≥2MΩ
5. Air supply: air pressure 0.55MPa(at connection box)
0.6MPa0.8MPa (compressed air),free from oil, water
6. Water supply: water pressure 0.2MPa0.4MPa
Recommend water hardness≤60mg/l
7. Fiber optic turbine handpiece
The handpiece will begin to work as soon as the handpiece is pulled and foot control is stepped
Slide foot control on left side (or right side), there are shown working function and working power on the LED
ASlide the foot control pedal on the left side, the handpiece is working with water. Adjusting the water volume valve could change the spray.
BSlide the foot control pedal on the light side, the handpiece is working without water. (Press the button - also could adjust the power)
CClean air switch of foot control pedal is stepped.the handpiece can spray clean air.
Adjust the illumination of fiber optic handpiece:pick up turbine handpiece, and press the fiber optic light button on the main control panel. Adjust the illumination by +and -.
Adjust water volume of fiber optic handpiece using the switch on the instrument tray to adjust water volume and spray.
8. High speed handpiece
Rotation≥300000r/min(4 holes)
Operation pressure: 0.25Mpa(according to the model)
9. Low speed handpiece
Pick up the low speed handpiece, adjust the speed and direction bySEL1andSEL2
a. Speed
When set the speed, adjust the rotation by+and -, and press SEL2 to confirm, then press SEL1 to return. (Automated return without operation in 3 seconds) (Fig.25)
10. Six way syringe
Take down the six way syringe from the instrument tray and push the corresponding button, then the water, air and spray can be obtained.
Warm water and warm spray are available by pressing the key twice.
11. Operating light
AC 17V 95W
12. Rinsing water heating button:
AC 24V 80W
Temperature scope: 20°-55°
13. Water pipe sterilization
Open the right door, pick up the sterilization bottom. Add 400ml disinfector (3% medical H2O2), and reset the bottom into the main chest, then close the door.
In the sterilization process, install the sterilization holder on the spittoon; keep pressing the sterilization button more than 4 seconds. The unit comes to the automated sterilization system. Insert the connecting pipe of handpiece into sterilization holder. The sterilization time will be 30 minutes. There is sign in LED , and close the sterilization bottom for stop the process immediately.
14. Suction ejector: air negative pressure
15. Saliva ejector: water negative pressure
Medical Instruments